Current Probes
Current probes required by various EMI specifications (such as MIL-STD-461/2) are
toroidal transformers designed to measure r.f. currents on active power lines or
other conductors.
Description
Direct connection to the conductor carrying EMI current is not necessary, since the
probe may be opened for insertion of the conductor into the window of the toroid
and then closed again to form a toroidal transformer with the conductor acting as
a one-turn primary.
A correction factor graph is provided to convert measured microvolts to EMI microamperes.When
the EMI current is measured in dB above one microvolt as indicated on a conventional
EMI meter, the correction factor will convert the measurement to dB above one microampere.
The correction factor is the inverse of the transfer impedance, Zt. Each probe is
shipped with a graph of the correction factor versus frequency, keyed to the serial
number on the probe.
Under certain conditions, a current probe can be used to inject low level r.f. signals
into individual wires or cable bundles. Ask our engineers for advice in the use
of current probes for this purpose. Some current probes can be easily damaged or
are otherwise unsuitable for this application.
Application
A current probe is used as a "pick-up" device for measuring r.f. current
in single conductors or cable bundles when connected to the 50 ohm input of a radio
frequency interference measuring receiver or spectrum analyzer.
Injection Probes
Specifications require the injection of large high frequency currents into cable
bundles and individual wires, using inserted secondary toroidal transformers placed
around the conductors being tested.
Description
Bulk Current Injection Probes are available in two styles:
1. Fixed window style where the wire(s) under test must be passed through the window.
2. A split toroidal design where the probe can be opened up and clamped over the
wire(s) under test.
Each probe is calibrated for insertion loss and transfer impedance in a test fixture
designed for the particular window size.This fixture provides a signal path with
a low Voltage Standing Wave Ratio. A typical fixture is Solar Type 9125-1,
used for probes with 32 to 44 mm diameter windows. Ask for details on this and other
test fixtures.
Injection probes available at the time of this printing are described later in this
section. Development of new styles is ongoing. If a probe meeting your requirements
is not found on the list, send us details and we will satisfy your need.
Application
High power r.f. amplifiers with 50 ohm output impedance are used to deliver voltage
to the injection probe. The wire or cable through the window of the probe acts as
a secondary of the toroidal transformers. This test method is intended to be used
instead of earlier methods, such as CS-01, CS-02, and RS-02 of MIL-STD-461.
Technical Information
Current probes used as receptors are known as �inserted primary
toroidal transformers� for connection to EMI receivers.
Injection probes which deliver high r.f. currents into wires through
the window are called �inserted secondary toroidal transformers�.
The maximum voltage carried on wires through the window is limited only by the insulation
of the wires. Maximum primary current in the wires through the window of current
probes is listed on the table below (symbol lp).
The signal input to injection probes is rated in watts from the signal source as
indicated in the table on the following page.
Development is continuing on new and useful probes, both current measuring �sensor�probes
and high wattage injection probes.The following is a partial list. As time goes
on, the list will grow. If you do not see what you need, just ask.
CLAMP-ON CURRENT PROBES
|
MAXIMUM PRI. (lp) CURRENT, AMPS |
Solar
Type No. |
Window
Diameter |
Nominal
ZT Ohmns |
DC to 60 Hz |
400 Hz |
RF (CW) |
PULSE |
FREQUENCY
RANGE |
| 9124-1N |
1.25" (32.0 mm) |
0.001 |
200 |
70 |
50 |
5000 |
1 KHz -200 MHz |
| 9204-1 |
1.25" (32.0 mm) |
0.33 |
350 |
350 |
50 |
100 |
1 KHz - 8 MHz |
| 9205-1 |
1.25" (32.0 mm) |
0.33 |
800 |
800 |
50 |
100 |
20 Hz - 8 MHz |
| 9118-1 |
1.25" (32.0 mm) |
0.10 |
350 |
150 |
22 |
500 |
500 Hz -200 MHz |
| 9134-1 |
1.25" (32.0 mm) |
0.70 |
500 |
400 |
5 |
100 |
20 Hz -100 MHz |
| 6741-1 |
1.25" (32.0 mm) |
0.70 |
350 |
225 |
5 |
100 |
10 KHz -100 MHz |
| 9206-1 |
1.25" (32.0 mm) |
1.0 |
350 |
350 |
4.2 |
100 |
10 KHz -150 MHz |
| 9207-1 |
1.25" (32.0 mm) |
1.0 |
800 |
450 |
4.2 |
100 |
20 Hz -150 MHz |
| 9208-1 |
1.25" (32.0 mm) |
1.0 |
350 |
350 |
100 |
200 |
10 KHz - 30 MHz |
| 9209-1 |
1.25" (32.0 mm) |
1.0 |
800 |
800 |
100 |
200 |
20 Hz - 30 MHz |
| 9145-1 |
1.25" (32.0 mm) |
5 |
350 |
350 |
42 |
100 |
10 KHz -150 MHz |
| 9119-1N |
1.25" (32.0 mm) |
1-7 |
200 |
200 |
40 |
70 |
1 MHz - 1.2 GHz |
| 9123-1N |
1.25" (32.0 mm) |
1-5 |
200 |
200 |
40 |
60 |
10 KHz -500 MHz |
| 9115-1N |
1.25" (32.0 mm) |
10 |
250 |
250 |
50 |
100 |
3 MHz-200 MHz |
| 9214-1 |
1.25" (32.0 mm) |
5 |
600 |
600 |
21 |
200 |
20 Hz -150 MHz |
| 9215-1N |
1.25" (32.0 mm) |
1-5 |
400 |
350 |
40 |
100 |
20 Hz -500 MHz |
| 9231-1 |
1.25" (32.0 mm) |
.010 |
300 |
200 |
7 |
3000 |
1 KHz -200 MHz |
| 9219-1N |
1.25" (32.0 mm) |
.025 |
400 |
300 |
150 |
200 |
20 Hz - 20 MHz |
| 9242-1 |
1.25" (32.0 mm) |
.025 |
400 |
80 |
35 |
100 |
20 Hz - 20 MHz |
| 9250-1N |
1.25" (32.0 mm) |
.10 |
200 |
200 |
10 |
200 |
10 KHz -450 MHz |
| 9136-1N |
2.62" (67.0 mm) |
5 |
350 |
350 |
100 |
200 |
10 KHz -100 MHz |
| 9249-1N |
2.62" (67.0 mm) |
8 |
350 |
350 |
60 |
200 |
10 KHz -300 MHz |
| 9256-1N |
2.62" (67.0 mm) |
2 |
350 |
350 |
200 |
200 |
10 KHz - 30 MHz |
| 9257-1N |
2.62" (67.0 mm) |
2 |
500 |
500 |
200 |
200 |
20 Hz - 30 MHz |
| 9258-1N |
2.62" (67.0 mm) |
5 |
500 |
500 |
100 |
200 |
20 Hz -100 MHz |
| 9260-1N |
2.62" (67.0 mm) |
2 |
350 |
75 |
2.6 |
100 |
10 KHz -200 MHz |
| 9261-1N |
2.62" (67.0 mm) |
2 |
500 |
500 |
2.6 |
100 |
20 Hz -200 MHz |
| 9262-1N |
2.62" (67.0 mm) |
.03 |
350 |
150 |
80 |
200 |
100 KHz - 100MHz |
| 9263-1N |
2.62" (67.0 mm) |
.3 |
500 |
500 |
80 |
200 |
20 Hz -100 MHz |
| 9301-1N |
2.62" (67.0 mm) |
8 |
500 |
500 |
60 |
200 |
20 Hz -500 MHz |
| 9302-1N |
2.62" (67.0 mm) |
.001 |
400 |
70 |
100 |
500 |
20 Hz -100 MHz |
| 9303-1N |
2.62" (67.0 mm) |
.001 |
500 |
200 |
100 |
5000 |
20 Hz -100 MHz |
| 9304-1N |
2.62" (67.0 mm) |
1 |
350 |
350 |
2.3 |
200 |
10 KHz -200 MHz |
| 9305-1N |
2.62" (67.0 mm) |
1 |
500 |
500 |
2.3 |
200 |
20 Hz -200 MHz |
| 9306-1N |
2.62" (67.0 mm) |
.005 |
200 |
65 |
60 |
5000 |
10 KHz -100 MHz |
| 9307-1N |
2.62" (67.0 mm) |
.005 |
300 |
300 |
60 |
5000 |
20 Hz -100 MHz |
| 9308-1N |
2.62" (67.0 mm) |
.01 |
300 |
75 |
17 |
2000 |
10 KHz -200 MHz |
| 9309-1N |
2.62" (67.0 mm) |
.01 |
450 |
450 |
17 |
2000 |
20 Hz -200 MHz |
FIXED WINDOW PROBES
|
MAXIMUM PRI. (lp) CURRENT, AMPS |
Solar
Type No. |
Window
Diameter |
Nominal
ZT OHMS |
DC to 60 Hz |
400 Hz |
RF (CW) |
Pulse |
Frequency
Range |
| 9218-1 |
.50" (12.7 mm) |
.1 |
10 |
10 |
8 |
100 |
10 KHz -250 MHz |
| 9248-1 |
.50" (12.7 mm) |
1 |
10 |
10 |
10 |
100 |
10 KHz - 50 MHz |
| 9203-1 |
.50" (12.7 mm) |
1 |
10 |
10 |
2.2 |
100 |
10 KHz -250 MHz |
| 9211-1 |
.50" (12.7 mm) |
.5-3 |
10 |
10 |
5 |
100 |
1 MHz- 1 GHz |
| 9202-1 |
.50" (12.7 mm) |
5 |
10 |
10 |
10 |
100 |
10 KHz -250 MHz |
| 9246-1 |
.50" (12.7 mm) |
.1 |
150 |
150 |
8 |
100 |
10 KHz -250 MHz |
| 9253-1 |
.50" (12.7 mm) |
1 |
150 |
150 |
65 |
100 |
20 Hz - 50 MHz |
| 9245-1 |
.50" (12.7 mm) |
1 |
150 |
150 |
3 |
100 |
10 KHz -300 MHz |
| 9210-1 |
.50" (12.7 mm) |
.3-5 |
150 |
150 |
16 |
100 |
300 KHz -500 MHz |
| 9212-1 |
.50" (12.7 mm) |
.3-4 |
150 |
150 |
30 |
100 |
1 MHz- 1 GHz |
| 9244-1 |
.50" (12.7 mm) |
5 |
150 |
150 |
13 |
100 |
10 KHz -600 MHz |
| 9346-1 |
.50" (12.7 mm) |
5 |
150 |
150 |
13 |
100 |
20 Hz -600 MHz |
CLAMP-ON INJECTION PROBES
Solar
Type No. |
Window
Diameter |
Winding
Current |
Rated
Watts |
Insertion Loss
Under 6 dB |
Insertion Loss
Under 10 dB |
Insertion Loss
Under 15 dB |
Insertion Loss
Under 20 dB |
Frequency
Range |
| 9108-1N |
1.25" (32 mm) |
10 |
50 |
- |
- |
120 KHz - 70 MHz |
60 KHz - 150 MHz |
10 KHz - 200 MHz |
| 9120-1N |
1.25" (32 mm) |
30 |
50 |
- |
12 MHz - 600 MHz |
7 MHz - 900 MHz |
4 MHz - 1GHz |
4 MHz - 1 GHz |
| 9121-1N |
1.25" (32 mm) |
30 |
50 |
- |
100 MHz - 550 MHz |
50 MHz - 800 MHz |
30 MHz - 1GHz |
10 MHz - 1 GHz |
| 9142-1N |
1.50" (38 mm) |
50 |
200 |
10 MHz - 350 MHz |
5 MHz - 430 MHz |
2.5 MHz - 500 MHz |
1.5 MHz -500 MHz |
2 MHz -500 MHz |
| 9144-1N |
1.50" (38 mm) |
26 |
100 |
- |
200 KHz - 8 MHz |
70 KHz - 90 MHz |
40 KHz -100 MHz |
10 KHz -100 MHz |
| 9217-1N |
1.50" (38 mm) |
26 |
100 |
800 KHz - 1.5 MHz |
500 KHz - 40 MHz |
250 KHz -100 MHz |
150 KHz -100 MHz |
10 KHz -100 MHz |
| 9310-1N |
2.62" (67 mm) |
26 |
100 |
- |
15 MHz - 450 MHz |
800 MHz - 650MHz |
4 MHz -800 MHz |
5 MHz -800 MHz |
| 9607-1N |
1.25" (32 mm) |
10 |
50 |
- |
500 KHz - 1 MHz |
200 KHz - 30 MHz |
120 KHz -300 MHz |
10 KHz -300 MHz |
Injection probes can also be used as current probes. A correction factor graph and
instructions for its use are supplied.
CALIBRATION FIXTURES
Note: Except for p/n 9125-1 and 9357-1, the probes are supported and centered in
the fixture.
Solar
Part No. |
DESCRIPTION |
| 9125-1 |
For Injection Probes, 1.5" (32 mm - 44 mm)
diameter window. 20 Hz to 500 MHz. Type BNC connectors. |
| 9125-2 |
For Solar 9119-1N Probe and similar, 1.25"
(32 mm) diameter window. 400 MHz to 3 GHz. Type N connectors. |
| 9251-1 |
For Eaton, Stoddart and Solar Probes, 1.25"
(32 mm) diameter window. 20 Hz to 500 MHz. Type BNC connectors. |
| 9254-1 |
For Eaton, Stoddart and Solar Probes, 2.62"
(66 mm) diameter window. 20 Hz to 500 MHz. Type N connectors. |
| 9321-1 |
For Eaton, Stoddart and Solar Probes, 0.75"
(19 mm) diameter window, totally enclosed. 20 Hz to 1.5 GHz. Type BNC connectors.
|
| 9330-1N |
For Eaton, Stoddart and Solar Probes, 1.25"
(32 mm) diameter window, totally enclosed. 20 Hz 1.0 GHz. Type N connectors.
|
| 9357-1 |
Clam Shell Fixture for Solar 9335-2 Probe.
20 Hz to 100 MHz. Type BNC connectors. |